HZKS-3 Automatic COC Cleveland Open Cup Flash Point Apparatus

Function Overview

Flash Point is the lowest temperature at which a liquid can form an ignitable mixture in air near the surface of the liquid. The lower the flash point, the easier it is to ignite the material. Nowadays, Flash point is used to say non-volatile or flammable substances whether exit highly volatile or flammable substances, and has been one of the most important parameters for petroleum products transportation, storage, operations, safety management.


Technical Parameters

Temperature Measurement

Temperature measuring range

-59.9°C~399.9°C

Temperature Control Range

Room temperature~400°C

Repeatability

≤150℃ 3℃, >150℃ 5℃

Resolution

0.1°C

Precision

0.5%

Temperature Sensor

Platinum Resistor(PT100)

Flash Fire Sensor

Ion detection ring

Ambient Temperature

0 ~ 40°C

Relative Humidity

≤ 85% RH

Power Supply

AC220V ± 20%; 50HZ ± 10%

Power Consumption

Approx 400W

Overall Dimension

440X280X290(mm)

Instrument Weight

14kg







Failure And Treatment

NO.

Failure

Treatment

1

Turn on the power, the LCD screen no display

Check power supply is plugged in, open the instrument measured the board examine the plug is not loose.

2

Large repeatability error for sample testing

Remove the lifting arm top cover, and check

1. whether the mixer shaft is broken

2. Oil may effect on thermocouple sensitivity, plz dry it by filter paper

3

No heating

Electric wire broken

4

Lift arm rise or fall failure

Limit opt coupler damage

The screw between main engine and lift arm fell off